The AirTherm™ SMT is a new generation of heat control system from
WPI designed to be used in Live Cell Imaging applications with micro-
scopes fitted with a full microscope environmental chamber enclosure.
The standard AirTherm™ SMT controls temperature and, as an option,
humidity.
The AirTherm™ SMT uses a PID control algorithm to maintain tightly
controlled loops of heat and humidity environment control.
With AirTherm™ SMT, the temperature of the sample and microscope
optics can be controlled to within 0.2°C. During operation, air is drawn out
of the chamber through a flexible hose, heated by the AirTherm™ SMT
heater and re-circulated to the chamber by the return hose.
Precision heat controller for use in live cell imaging and custom
incubators
Control heat and humidity (optional) with a single controller
Electrically and acoustically quiet
Quick, precise response to thermal change
The system is typically used in a closed loop configuration.
The AirTherm™ SMT system includes:
Two coil-reinforced heater hose pieces and hose clamps.
Temperature sensor for remote placement in environmental chamber
Humidity probe for monitoring chamber humidity available as an option.
Precision Heat Control
Smart, electrically quiet air heater for live cell
imaging systems and custom incubators
AIRTHERM SPECIFICATIONS
AIR FLOW RATE 20–50 CFM (0.55–1.4 m
3
/minute)
CONTROL TEMPERATURE RANGE Ambient to 60°C
TEMPERATURE RESOLUTION 0.1°C
TEMPERATURE ACCURACY 0.2°C
HUMIDITY CONTROL RANGE Ambient–90%
HUMIDITY RESOLUTION 0.1%
HUMIDITY ACCURACY 5%
HUMIDITY REPEATABILITY 0.5%
ANALOG OUTPUT 0.5°C resolution;
FOR CHART RE-CORDER 0–10V represents 0–100°C
HEATING VOLUME Less than 50 CF (1400L), re-circulating
TEMPERATURE SENSOR TYPE Platinum RTD 1000 W
HUMIDIFIER TYPE Ultrasonic
HUMIDIFIER TANK CAPACITY 0.5 gallons
HUMIDIFIER DAILY OUTPUT 2 gallons
FUSE For 120 VAC, 8A 250 V 5x20 mm metric
For 230 VAC, 4A 250 V 5x20 mm metric
POWER 450 W, 95–135 V or 220–240 V, 50/60 Hz
DIMENSIONS 6½ x 8 x 7½ in. (15.5 x 21 x 19 cm)
Microscope
Chamber
(not included)
Air-Therm SMT
A typical AirTherm™ SMT installation
places the heated air inflow at the
bottom and the cold air return at the
top of the microscope chamber.
World Precision Instruments www.wpiinc.com
UK: Tel: +44 (0)1462 424700 • [email protected] Germany: Tel: +49 (0)30-6188845 • [email protected]
US: Tel: 941-371-1003 • sales@wpiinc.com
AIRTHERM-SMT-1W Airtherm™ SMT Heater, 110V
AIRTHERM-SMT-2W Airtherm™ SMT Heater, 230V
OPTIONAL ACCESSORIES
15590 Replacement Hoses, 2.5” diam., 4.5 ft
300276 Replacement Platinum Temperature Probe
98727 Replacement Temperature Probe
98728 Humidity Probe