4 World Precision Instruments
Setup
1. Remove the microscope from the case and that the parts are all present and
undamaged. See “Parts List” on page 2.
2. Assemble the track stand.
3. Mount the microscope head into the focus mount.
4. Remove the protective cap on the bottom of the microscope head.
5. Refer to the parts diagram (Fig. 2) and follow the set-up instructions.
6. Install the WFH 10x eyepiece pair (included). The paired eyeguards can now be
attached to the eyepieces.
7. The exact working distance is found by adjusting the focusing knob.
8. When the optional stands are used instead of the standard base, see “System
Diagram” on page 5 for mounting instructions.
Adusting a Microscope
These instructions describe how to properly focus a binocular microscope.
1. Turn on the lamp and adjust its intensity.
2. Adjust the interpupilary distance so that the two circles of light merge into one.
3. Place a sample on the stage.
4. Select the 10X objective and look through the right eyepiece only.
5. Use the coarse and fine focus control knobs to adjust the focus of the sample.
6. When the sample is clearly visible, use only your left eye. Do NOT adjust the
focus knobs. Instead, adjust the diopter on the left eyepiece until the sample
comes clearly into view. The image should be clear with both eyes open.
7. Choose the desired objective. You may need to adjust the fine focus a little.
TIP: For information on microscope basics, understanding objectives, choosing a
microscope camera, a camera comparison chart and a camera/microscope
configuration chart, see www.wpiinc.com/microscope.